pH meter level and accuracy
The pH meter's level and its accuracy are two distinct concepts. The "level" of a pH meter refers to the resolution or the smallest display value of the instrument, while accuracy is the overall measurement error that includes both the meter and the electrode. For example, a meter with a 0.1 pH division is classified as a 0.1-level instrument, and one with a 0.001 pH display is called a 0.001-level instrument.
However, the accuracy of the pH meter is not solely determined by the meter itself. It also depends on the performance of the glass and reference electrodes. Even though modern technology allows for highly precise meters with resolutions as fine as 0.001 pH, the accuracy is still limited by the electrodes. Glass electrodes can have repeatability errors of up to 0.01 pH, and reference electrodes may suffer from junction potential instability. As a result, even with high-resolution meters, the overall test accuracy rarely exceeds 0.01 pH.
Choosing a higher-resolution meter helps reduce the impact of the meter’s limitations on measurement errors, but it doesn’t guarantee perfect accuracy. The real challenge lies in improving the reliability, usability, and cost-effectiveness of pH meters. Ongoing research continues to enhance the intelligence and user-friendliness of these instruments.
According to the national verification regulations (JJG 119-84), the relationship between the level of a pH meter and its test accuracy is clearly defined. For instance:
| Level | Division Value (pH) | Electrometer Error (pH) | Total Indication Error (pH) |
|-------|---------------------|--------------------------|-------------------------------|
| 0.2 | 0.2 | ±0.1 | ±0.2 |
| 0.1 | 0.1 | ±0.05 | ±0.1 |
| 0.02 | 0.02 | ±0.01 | ±0.02 |
| 0.01 | 0.01 | ±0.01 | ±0.02 |
| 0.001 | 0.001 | ±0.002 | ±0.01 |
It's important to note that the total error for a 0.01-level meter is 0.02 pH, and for a 0.001-level meter, it's only ±0.01 pH when using primary standard materials. Most commonly used pH meters rely on secondary standards, which are less precise.
Additionally, the test range must be controlled between pH 3 and pH 10. This is because the liquid junction potential of the reference electrode becomes unstable outside this range, leading to larger errors—up to ±0.07 pH. Moreover, glass electrodes can introduce alkali or acid errors in extreme pH conditions. Therefore, specifying the test range ensures accurate comparison of different pH meters and avoids confusion caused by external factors.
We can provide rubber handball and machine stitched handball ball.Rubber surface is waterproof,improve abrasion resistance and skid resistance.Out of the rubber bladder is nylon yarn,which can maintain long-lasting roundness and make handball durable to play.We can also provide different leather materials for handball shells,like PVC,PU,TPU.
JYMINGDE assists you throughout the project: ball sizes, color, logo, design and packaging. Contact us or mark your requirements on enquiry quote.
Size | Weight(g) | Diameter(inch) | Packing(pcs/ctn) | Carton Size(cm) |
G.W./N.W.(kg) |
Size 3 Handball | 400-430 | 7.27-7.5 | 50 | 47*42*33 | 25/23 |
Size 2 Handball | 320-350 | 6.77-7 | 50 | 47*38*31 | 20/18 |
Size 1 Handball | 270-300 | 6.27-6.5 | 50 | 43*38*30 | 17/15 |
Handball,Hand Ball,Handball Ball,Handballs
Jiangyin Mingde Sports Goods Co., Ltd. , https://www.mdsportsball.com